Last updated: November 2026
Linked FDA sources and CFR provisions: page opened November 26, 2026; CFR text checked against the GPO eCFR file built November 13, 2026.
The FDA premarket pathway for software as a medical device is the set of decisions a software maker works through before selling: whether each software function is a device, whether premarket review is required and which route applies, and what quality system and documentation stand behind the submission. Integral Healthcare Solutions (IHS) drafts the pathway worksheet, documentation outline and procedures for the manufacturer's regulatory lead to approve.
What is the FDA premarket pathway for software as a medical device?
FDA, using the International Medical Device Regulators Forum definition, calls software as a medical device software intended for one or more medical purposes that performs them without being part of a hardware medical device (FDA SaMD page, content dated December 4, 2018, page opened November 26, 2026). FDA says its policies are function-specific and apply across platforms (FDA device software functions page, page opened November 26, 2026).
- Is the function a device? The device definition does not include software functions excluded under section 520(o) of the Federal Food, Drug, and Cosmetic Act (FDA device determination page). For clinical decision support aimed at health care professionals, FDA's guidance issued January 29, 2026, which supersedes a January 6, 2026 version, says a function is excluded only if it meets all four criteria in section 520(o)(1)(E), which turn on the kind of input data, what the software does with medical information, whom it advises and whether that professional can independently review the basis without relying primarily on the recommendations for an individual patient’s diagnosis or treatment (Clinical Decision Support Software (January 29, 2026)). Existing digital health policies still apply to device functions meant for patients or caregivers (FDA guidance page). FDA's Digital Health Policy Navigator helps with this step, but its results are not a formal FDA device determination (Navigator page).
- Which route? This page compares 510(k), De Novo and premarket approval; exemptions and other pathways may apply.
- What quality system stands behind the submission? FDA's Quality Management System Regulation (QMSR), 21 CFR Part 820, became effective February 2, 2026 and incorporates ISO 13485:2016 by reference (FDA QMSR page). Under section 820.10(c)(1), manufacturers of class I devices automated with computer software must meet the design and development requirements in clause 7.3 of that standard (21 CFR 820.10).
Who needs it and what triggers it?
A software maker needs a pathway decision whenever a function may meet the device definition. For many device functions that pose minimal risk, FDA says it will exercise enforcement discretion and will not expect premarket review applications or registration and listing (FDA device software functions page). A device outside that discretion may still be exempt from premarket notification, subject to the exemption’s limitations.
How does IHS help with the premarket pathway?
IHS works under your regulatory or quality lead in its Compliance Services practice line. We assess gaps, map evidence, draft content, conduct a mock review and support readiness. Your regulatory lead approves the materials, and your clinicians approve clinical content.
- A function-by-function pathway worksheet written up as the manufacturer's own assessment for your lead to review and sign.
- A route comparison memo against FDA's stated conditions, for your regulatory lead and counsel to decide on.
- A documentation outline mapped to FDA's software content guidance, with a gap list.
- Draft question lists for any 513(g) request or Pre-Submission, for your team to finalize.
- Draft software-lifecycle procedures aligned to Part 820 (document control: our FDA QMSR document control page).
You supply the product facts, intended use statement, risk analysis, verification and validation records, any clinical data, and the eSTAR entries.
The real limit: pathway assessment documented as the client's; any 513(g) or pre-sub goes from the manufacturer; no clearance claim. IHS is a consulting firm, not a law firm, and does not test software.
What should you have ready?
- A written intended use and indications for use for each software function. FDA tells sponsors to define these first when deciding whether a product is a device (FDA device determination page).
- A function inventory with Navigator results (Navigator page).
- For decision-support functions, a written test against all four section 520(o)(1)(E) criteria (Clinical Decision Support Software (January 29, 2026)).
- A predicate comparison under 21 CFR 807.100(b) (807.100) or a De Novo classification summary under 21 CFR 860.220(a)(8): searches and rationale if there was no prior 510(k), or the NSE 510(k) number and a search for subsequent classification. Section 860.220(c) permits justified omissions (860.220).
- A Documentation Level statement and rationale. Assess software failures or flaws before risk controls, in the intended-use context: FDA recommends Enhanced for probable death or serious-injury risks to patients, users or others, excluding purely hypothetical risks; Basic applies otherwise. Check the guidance’s additional device-category recommendations (Content of Premarket Submissions for Device Software Functions (June 14, 2023)).
- A risk management file, which FDA recommends providing with the submission (Content of Premarket Submissions for Device Software Functions (June 14, 2023)).
- A documented quality management system meeting applicable ISO 13485:2016 and Part 820 requirements for the manufacturer’s operations, including applicable design and development records (21 CFR 820.10).
- For 510(k) devices, a change assessment under 21 CFR 807.81(a)(3) and its exceptions; for AI-enabled functions, if proposing a PCCP, a draft covering planned modifications, the method to develop, validate and implement them, and an impact assessment (Marketing Submission Recommendations for a Predetermined Change Control Plan for Artificial Intelligence-Enabled Device Software Functions (August 18, 2025)).
- A list of prior FDA submissions for this device, including 513(g) requests and pre-submissions, or a statement that there were none; justified omissions are addressed in 21 CFR 860.220(c) (21 CFR 860.220).
Bring this list to the introductory call.
How do the routes compare?
| Route | FDA's stated use | Before marketing | FDA timing statement |
|---|---|---|---|
| 510(k) | Show substantial equivalence to an eligible legally marketed predicate: same intended use and either the same technological characteristics, or different characteristics with data demonstrating equivalent safety and effectiveness and no different safety/effectiveness questions. Clinical data are required if FDA deems them necessary. The predicate must satisfy 21 CFR 807.100(b)(3) (510(k) page) | The submitter may not market until it receives an order declaring substantial equivalence | Determination usually within 90 days |
| De Novo | Classify a novel device with no legally marketed predicate into class I or II, where general controls, or general and special controls, give reasonable assurance of safety and effectiveness (De Novo page) | A granted device may be marketed and used as a predicate for later 510(k)s, when applicable | Review days exclude time on hold for an Additional Information request |
| Premarket approval | Class III devices subject to PMA under 21 CFR Part 814; some preamendment class III devices may use 510(k). FDA evaluates valid scientific evidence of safety and effectiveness (PMA page) | The applicant must receive approval before marketing | No timing statement on the page we read |
A De Novo request can follow a high-level NSE finding (no predicate, new intended use, or different technological characteristics raising different safety/effectiveness questions), or proceed directly when no legally marketed device can support substantial equivalence. FDA must receive a complete response to every deficiency within 180 calendar days of the Additional Information letter’s date; otherwise it considers the request withdrawn and deletes it from review. No extensions are granted (De Novo page), and FDA may decline a request for a device of a type already classified (21 CFR 860.260).
Two FDA interactions are not routes. A 513(g) request for information asks FDA for information on a device's classification and requirements, and FDA says its response is not a classification decision and is not clearance or approval (513(g) guidance). A Pre-Submission is one type of Q-Submission: it requests written FDA feedback, optionally followed by a meeting (Q-Submission guidance). FDA recommends considering one before a De Novo request (De Novo page).
What does it cost?
FDA publishes its device user fees, and for fiscal year 2027 (October 1, 2026 through September 30, 2027) the standard fees and fees for businesses qualified by CDRH as small businesses are: 510(k), 28,653 dollars and 7,163 dollars; De Novo request, 191,020 dollars and 47,755 dollars; premarket approval application, 636,732 dollars and 159,183 dollars; 513(g) request, 8,596 dollars and 4,298 dollars (FDA MDUFA fees page). A 510(k) submitted to FDA on behalf of an FDA-accredited third-party reviewer carries no FDA user fee. Verify current fees with FDA.
IHS scopes each engagement after a free introductory call.
What this is not
- Not legal advice. IHS is a consulting firm, not a law firm.
- Not an FDA determination, clearance or approval, and not a prediction of one.
- Not a filing service. The pathway assessment stays the manufacturer's, and any 513(g) request, Pre-Submission or premarket submission goes from the manufacturer.
Frequently asked questions
What is software as a medical device?
FDA uses the International Medical Device Regulators Forum definition: software intended for one or more medical purposes that performs them without being part of a hardware medical device. FDA's policies are function-specific.
Is my clinical decision support software a device?
The CDS exclusion in section 520(o)(1)(E) requires all four criteria; other statutory exclusions may apply. Functions intended to acquire, process or analyze medical images, IVD signals, or patterns or signals from a signal acquisition system remain devices when intended for a purpose in the statutory device definition.
Does FDA regulate every health app?
No. For many device functions that pose minimal risk, FDA says it will exercise enforcement discretion and will not expect premarket review applications or registration and listing. Its examples include functions that help patients self-manage a condition without specific treatment suggestions.
What is the difference between a 510(k) and a De Novo request?
A 510(k) shows substantial equivalence to an eligible legally marketed predicate. De Novo classifies a novel device without a predicate into class I or II when general controls, or general and special controls, provide reasonable assurance of safety and effectiveness; granted devices may serve as predicates, when applicable.
When is premarket approval needed?
PMA generally applies to class III devices, although some preamendment class III devices may use 510(k). When PMA is required, FDA approval must precede marketing.
Does a 513(g) response clear my software?
No. FDA says its responses to 513(g) requests are not device classification decisions and do not constitute FDA clearance or approval for marketing.
Can I change my software after clearance without a new submission?
A new 510(k) is generally required for a change that could significantly affect safety or effectiveness, or a major intended-use change, subject to 21 CFR 807.81(b). A change consistent with a PCCP cleared under section 515C does not require a new 510(k).
What quality system does a software device maker need?
Manufacturers subject to Part 820 must maintain an appropriate quality management system for their devices and operations. The QMSR took effect February 2, 2026 and incorporates ISO 13485:2016; section 820.10(c) requires manufacturers of class II, class III and specified class I devices, including those automated with software, to meet clause 7.3 and its subclauses.
What does IHS do, and what does the manufacturer do?
Integral Healthcare Solutions drafts the pathway worksheet, route comparison memo, documentation outline, question lists and procedures from facts your team supplies. Your regulatory lead decides the pathway, and the manufacturer sends any 513(g) request, Pre-Submission or submission to FDA. IHS makes no clearance claim.