FDA Premarket Cybersecurity Documentation Support for Medical Device Makers (Section 524B)

Last updated: November 2026

Section 524B governs cybersecurity duties for covered medical-device submissions. IHS drafts documentation from the manufacturer’s engineering records for its team to approve; the manufacturer designs and tests device security and submits to FDA.

What is FDA premarket cybersecurity documentation under section 524B?

It is the information a sponsor must include in a premarket submission for a cyber device, set by section 524B of that Act, enacted December 29, 2022 as part of Pub. L. 117-328 (21 U.S.C. 360n-2 at GPO, page opened November 26, 2026). The statute names three duties. First, submit a plan to monitor, identify and address, as appropriate and in a reasonable time, postmarket cybersecurity vulnerabilities and exploits, including coordinated vulnerability disclosure and related procedures. Second, design, develop and maintain processes and procedures providing reasonable assurance that the device and related systems are cybersecure, and make updates and patches available to both: on a reasonably justified regular cycle for known unacceptable vulnerabilities, and as soon as possible out of cycle for critical vulnerabilities that could cause uncontrolled risks. Third, an SBOM covering commercial, open-source and off-the-shelf software components. A fourth clause lets FDA require more by regulation.

FDA's guidance, Cybersecurity in Medical Devices: Quality Management System Considerations and Content of Premarket Submissions, was issued February 3, 2026 and supersedes the June 27, 2025 version. Section VII of the guidance addresses section 524B (FDA guidance, page opened November 26, 2026; FDA guidance page).

What counts as a cyber device?

A cyber device meets all three statutory tests: software validated, installed or authorized by the sponsor as a device or in a device; ability to connect to the internet; and technological characteristics validated, installed or authorized by the sponsor that could be vulnerable to cybersecurity threats (statute). FDA treats firmware and programmable logic as software, and the ability to connect as present whether or not the connection is intended. Its illustrative list includes network, server and cloud connections, radio-frequency links such as Wi-Fi, cellular and Bluetooth, magnetic inductive communication, and hardware connectors capable of connecting to the internet, such as USB, ethernet and serial ports. FDA's example is a device serviced over a brief USB connection, which still has the ability to connect (guidance Section VII.B).

What is required and what is only recommended?

The three duties above are statutory. The documentation detail in the guidance is not: FDA says guidances describe its current thinking and are recommendations unless specific regulatory or statutory requirements are cited, and that should means suggested, not required (guidance, Introduction). FDA says documentation breadth scales with cybersecurity risk, but each type of documentation is recommended for every submission for a device with potential cybersecurity risks (guidance Appendix 4).

Who needs it and what triggers it?

Section 524B covers persons submitting 510(k)s, PMAs, PDPs, De Novo requests or HDEs for cyber devices, including Abbreviated/Special 510(k)s and PMA/HDE supplements, subject to any FDA exemption under subsection (d). FDA says the requirements do not apply to submissions made before March 29, 2023 (FDA cybersecurity FAQ, page opened November 26, 2026). The following submission situations and guidance updates help frame preparation; a guidance revision or screening hold does not independently trigger section 524B.

How does IHS help with a section 524B submission?

IHS drafts documentation, led by Tom (Thomas G. Goddard, JD, PhD) and built from the manufacturer's own engineering and quality records, in five steps.

  1. Gap assessment: compare records with statutory duties and FDA’s Table 1.
  2. Draft the management plan, disclosure procedure, patch-cycle justification and reasonable-assurance procedures for manufacturer approval.
  3. Map SBOM data requirements and draft justifications; engineers generate the SBOM.
  4. Map engineering evidence to controls and tests; draft labeling.
  5. Mock review: check the assembled package and identify gaps for the manufacturer to resolve.

What the manufacturer supplies: the device architecture, threat modeling and security design decisions, all security testing, the component list, the eSTAR entries, the submission, and every contact with FDA. The real limit: IHS performs no penetration, vulnerability or other security testing, makes no statement that a device is secure, and promises no FDA outcome.

What should you have ready before you submit?

This checklist distinguishes statutory duties from recommendations in FDA’s Cybersecurity in Medical Devices: Quality Management System Considerations and Content of Premarket Submissions (February 3, 2026), page opened November 26, 2026. FDA describes Table 1 as an organizational aid, not merely a deliverable checklist.

  1. IHS preparation item: record the three-part cyber-device assessment (section 524B(c)).
  2. The required postmarket vulnerability plan; FDA recommends nine management-plan elements (guidance VI.B and VII.C.1).
  3. Disclosure procedures addressing vulnerabilities and exploits identified externally or by the manufacturer (guidance VII.C.1).
  4. Updates and patches for the device and related systems: a reasonably justified regular cycle for known unacceptable vulnerabilities; as soon as possible out of cycle for critical vulnerabilities that could cause uncontrolled risks (section 524B(b)(2)).
  5. Evidence of processes and procedures designed, developed and maintained to provide reasonable assurance of cybersecurity for the device and related systems (section 524B(b)(2)).
  6. The required SBOM; FDA recommends machine-readable baseline attributes, plus component support levels and end-of-support dates within it or separately, with justification for unavailable information (guidance V.A.4.b and VII.C.3).
  7. Known vulnerabilities, including CISA catalog entries, discovery methods, safety/security risk assessments and applicable risk controls (guidance V.A.4.b).
  8. Threat model, risk assessment and architecture views covering the global system, multi-patient harm, updateability/patchability and safety-related security use cases (guidance V.A, V.B and Appendix 4).
  9. Security testing documentation and associated reports or assessments (guidance V.C).
  10. Cybersecurity labeling, including relevant network ports and interfaces, their functions, direction and approved endpoints (guidance VI.A).
  11. For modifications unlikely to affect cybersecurity, the plan/reference, vulnerability/remediation information and SBOM described in the modification FAQ below (guidance VII.D.2).

Tom reviews this list with you on the introductory call.

How does section 524B documentation compare with the other FDA cybersecurity and quality documents?

Section 524B documentation is premarket submission content. These documents sit beside it.

Document or ruleWhat it coversStatus
Section 524B, Food, Drug, and Cosmetic ActPostmarket plan, reasonable-assurance processes and SBOM for cyber devicesStatute; listed-pathway cyber-device submissions from March 29, 2023, subject to any FDA exemption under subsection (d)
FDA premarket cybersecurity guidance (February 3, 2026)FDA's recommended design, labeling and documentation; Section VII covers 524BRecommendations, except for cited statutory or regulatory requirements
FDA postmarket cybersecurity guidance (December 27, 2016)Structured management of vulnerabilities in marketed devicesRecommendations (FDA FAQ listing)
QMSR, 21 CFR Part 820The manufacturer's quality management system; effective February 2, 2026Regulation; see our page on QMSR document control

Related IHS pages: device establishment registration and listing and medical device reporting and complaint handling.

What does FDA premarket cybersecurity documentation cost?

The FDA pages reviewed do not list a separate cybersecurity-documentation fee. This does not establish that the underlying submission is fee-free; verify current fees with FDA. IHS scopes each engagement after a free introductory call.

What this is not

Frequently asked questions

What does section 524B require of a device maker?

Sponsors must submit a plan addressing postmarket cybersecurity vulnerabilities and exploits through monitoring, identification and appropriate action in a reasonable time, including coordinated vulnerability disclosure and related procedures; design, develop and maintain processes and procedures providing reasonable assurance that the device and related systems are cybersecure; and provide an SBOM covering commercial, open-source and off-the-shelf software components. They must make updates and patches available to the device and related systems on a reasonably justified regular cycle for known unacceptable vulnerabilities, and as soon as possible out of cycle for critical vulnerabilities that could cause uncontrolled risks. They must also comply with additional requirements FDA establishes by regulation.

What is a cyber device?

A cyber device includes software validated, installed or authorized by the sponsor as a device or in a device; can connect to the internet; and contains technological characteristics validated, installed or authorized by the sponsor that could be vulnerable to cybersecurity threats. All three conditions must be met.

Does a USB service port make my device a cyber device?

It can, if all three cyber-device conditions are met. FDA includes hardware connectors capable of connecting to the internet, such as USB, ethernet and serial ports; its example counts a brief USB service connection toward the connectivity condition.

Is the FDA premarket cybersecurity guidance mandatory?

The guidance itself is not. FDA says its guidances are recommendations unless specific regulatory or statutory requirements are cited. Section 524B’s duties are binding for covered submissions, subject to any FDA exemption under subsection (d).

When did the requirements start, and what happened to the refuse-to-accept policy?

FDA says the requirements do not apply to submissions made before March 29, 2023. FDA's temporary refuse-to-accept transition policy expired on October 1, 2023; FDA then expected sponsors to have had sufficient time to prepare submissions meeting section 524B.

What goes in a cybersecurity management plan?

FDA recommends nine elements, among them the personnel responsible, how and how often vulnerabilities are monitored, a patch timeline, update processes, the coordinated vulnerability disclosure process, and how remediations are communicated to customers. FDA recommends submitting the plan with the premarket submission.

What does FDA recommend an SBOM contain?

The statute requires an SBOM covering commercial, open-source and off-the-shelf components. FDA recommends a machine-readable SBOM that follows the NTIA baseline attributes, plus each component's support level and end-of-support date, and a written justification for anything the manufacturer cannot provide.

Do I need this documentation again when I change a marketed device?

For a cyber device modification requiring a new submission under a listed pathway, section 524B applies. If cybersecurity is unlikely to be affected, FDA recommends referencing a previously submitted plan and summarizing changes; if none was submitted, a plan is required. FDA permits reduced process documentation describing current critical vulnerabilities that could cause uncontrolled risks and whether vulnerabilities with uncontrolled risk were remediated since the last authorization, including how. An SBOM is still required.

What does a consultant do here, and what does the manufacturer do?

Integral Healthcare Solutions drafts the management plan, disclosure and patching procedures, SBOM justification language, traceability map and labeling text from facts your engineers supply. Your team designs the device security, runs all security testing, generates the SBOM, makes the eSTAR entries and submits to FDA. IHS does not contact FDA or promise an FDA outcome.