Last updated: November 2026
Linked FDA pages, the FDA guidance PDF and the U.S. Code section were opened November 26, 2026. The statute text is the 2024 U.S. Code edition at govinfo.
A clinical decision support (CDS) device-status assessment examines each software function against FDA policy, including the section 520(o)(1)(E) exclusion. Integral Healthcare Solutions (IHS) drafts the manufacturer’s assessment for its regulatory lead and counsel to approve; this is not an FDA determination.
What is a CDS device-status assessment?
It records the manufacturer’s analysis of each function. FDA’s Clinical Decision Support Software guidance (January 29, 2026) explains the CDS exclusion. The guidance is nonbinding; alternative approaches must satisfy applicable statutes and regulations.
The current guidance was issued January 29, 2026 and supersedes a version issued January 6, 2026 (same source; FDA guidance page). It covers software for health care professionals. It does not address which other FDA requirements apply to a function that is a device (same source).
What are the four criteria FDA applies?
Exclusion under section 520(o)(1)(E) requires all four criteria below; other exclusions may apply (21 U.S.C. 360j(o); FDA guidance). Sections 520(o)(3)–(4) preserve additional authority, including over blood-manufacturing/transfusion software and software meeting section 513(a)(1)(C). In IHS’s words:
- No intended image/signal analysis. The function is not intended to acquire, process or analyze medical images, signals from in vitro diagnostic devices, or patterns or signals from signal acquisition systems. FDA’s “pattern” includes multiple, sequential or repeated signal measurements; discrete point-in-time physiological measurements generally are not patterns by themselves.
- Medical information. The function is intended to display, analyze or print patient or other medical information. FDA distinguishes a single clinically meaningful test result from continuous sampling, such as continuous glucose monitor readings.
- Recommendations for a professional. The function is intended to support or recommend prevention, diagnosis or treatment of a disease or condition to a health care professional. FDA says a specific preventive, diagnostic or treatment output or directive fails criterion 3. Lists of preventive, diagnostic or treatment options, including prioritized lists, and lists of follow-up or next-step options can meet criterion 3 if not intended to replace or direct professional judgment.
- Independent review. The function is intended to enable the professional to independently review the recommendations’ basis, so the professional is not intended to rely primarily on them for an individual patient’s clinical diagnosis or treatment decision.
Software that supports patients or caregivers rather than health care professionals meets the device definition, per the guidance, and FDA's existing digital health policies continue to apply to it (guidance page).
What do the edge cases in the current guidance say?
One clinically appropriate recommendation. If only one option is clinically appropriate and the function otherwise meets all the criteria, FDA intends to exercise enforcement discretion, meaning it does not intend to enforce requirements under the Act for that function (guidance). The guidance lists variations outside this policy, such as a cardiovascular risk estimate for the next 24 hours.
Time-critical use. FDA does not consider functions intended for a critical, time-sensitive task or decision to meet criterion 4, because the professional is unlikely to have time to review the basis (guidance). Its examples of device functions include one that detects a life-threatening condition such as stroke or sepsis and generates an alarm or alert. FDA's FAQ adds that being used in a setting such as the emergency department does not by itself make a product ineligible (CDS FAQs, content current June 29, 2026).
Failing a criterion is not the end of the analysis. FDA says many functions that fail one or more criteria are still not devices, or are device functions for which it intends to exercise enforcement discretion, and that the CDS guidance should not be the sole reference (CDS FAQs).
Multiple functions. A product can contain non-device CDS functions and device functions. The statute bars FDA from regulating the excluded function as a device, while allowing FDA to assess its impact when reviewing the device functions (21 U.S.C. 360j(o)(2)).
Reserved authority. For functions under section 520(o)(1)(C), (D) or (E), the Secretary may issue a final order after finding reasonably likely serious adverse health consequences. The Secretary must publish a notification and proposed order in the Federal Register. The notification must state the finding and rationale and identify the supporting evidence; at least 30 calendar days must be allowed for public comment before a final order or withdrawal of the proposed order. Required considerations include the likelihood and severity of patient harm if the function fails to perform as intended, its intended support for a health care professional’s clinical judgment, the professional’s reasonable opportunity to review the basis of its information or treatment recommendation, and the intended user and user environment (21 U.S.C. 360j(o)(3)).
Who is this assessment support for?
IHS offers this service to EHR, workflow and digital health software makers evaluating decision-support functions. FDA considers developers providing access to device software through subscriptions or software as a service to be device manufacturers (CDS FAQs).
IHS’s process recommendation, November 26, 2026: review a new function, changed inputs or outputs, or revised guidance. These are assessment checkpoints, not FDA filing requirements. FDA’s Digital Health Policy Navigator assesses functions individually.
How does IHS help?
IHS works under the manufacturer's regulatory or quality lead in its Compliance Services practice line. We assess gaps, map evidence, draft content, and run a mock review of the written assessment. The deliverables are drafts for your regulatory lead and counsel to approve:
- A function inventory that states each function's intended use, user and patient population.
- A criterion-by-criterion worksheet for each function, written as your organization's own assessment against the statute and FDA's guidance, with the evidence behind each answer.
- A criterion 4 gap list comparing your labeling and screens with the transparency items FDA recommends.
- A change trigger procedure so a new input or output prompts a fresh assessment.
- Draft question lists for your own use if you decide to ask FDA for feedback.
You supply the product facts, intended use statements, algorithm and validation information, labeling, screenshots and any usability results. Your regulatory lead decides the conclusion. Any inquiry to FDA is yours: the manufacturer sends it, and IHS does not contact FDA. IHS is a consulting firm, not a law firm, does not test software, and gives no legal advice.
What should you have ready?
Bring these materials for IHS’s documentation review; guidance recommendations are not a separate filing requirement.
- Intended use and indications for each function (FDA device determination page, opened November 26, 2026).
- A function inventory (Digital Health Policy Navigator, opened November 26, 2026).
- Input data types for criteria 1–2 (CDS guidance, January 29, 2026).
- Output types and clinical purpose for criterion 3 (CDS guidance, January 29, 2026).
- Intended professional users, patient populations and time sensitivity, criterion 4 (CDS guidance, January 29, 2026).
- Recommendation sources, including study or guideline versions/dates, criterion 4 (CDS guidance, January 29, 2026).
- Algorithm development, data and validation results, criterion 4 (CDS guidance, January 29, 2026).
- Screens/labeling with input instructions, relevance, data quality, patient-specific information and knowns/unknowns, criterion 4 (CDS guidance, January 29, 2026).
- Available usability evidence; testing may be needed in some cases, criterion 4 (CDS guidance, January 29, 2026).
- Other potentially applicable policies (CDS FAQs, June 29, 2026).
Bring this list to the introductory call.
How do the ways of getting an answer compare?
IHS names FDA's routes without recommending one.
| Route | What FDA says |
|---|---|
| Documented self-assessment | The statute and guidance set the test; FDA says the guidance is one of several references (FAQs). |
| Digital Health Policy Navigator | A seven-step tool run for each software function; its results are not a formal FDA device determination (Navigator). |
| Digital Health inbox or device determination mailbox | FDA lists a digital health questions inbox. Its device determination mailbox is for unresolved questions after reviewing its device determination page and contacting DICE (FAQs; device determination page). |
| 513(g) request for information | FDA points to a 513(g) request for a formal device determination or classification; its responses are not classification decisions, not clearance or approval, and not final Agency action (device determination page; 513(g) guidance, August 2024). |
| Pre-Submission | FDA suggests considering the Pre-Submission process for more specific feedback about a particular device (FAQs). |
What does it cost?
This page does not quote FDA fees; verify any applicable fee with FDA for the route under consideration. IHS scopes each engagement after a free introductory call.
What this is not
- It is not an FDA determination, clearance or approval, and it does not predict one.
- It is not legal advice. IHS is a consulting firm, not a law firm; the manufacturer and its counsel decide each function’s status and send any FDA inquiry. IHS does not contact FDA.
- It does not cover the requirements that apply once a function is a device, such as premarket submissions or the quality system. See the related pages below.
Frequently asked questions
Is clinical decision support software a medical device?
Not always: section 520(o)(1)(E) requires all four criteria for its CDS exclusion, subject to the reserved authority in sections 520(o)(3)–(4). Other exclusions and FDA policies may apply.
What are the four non-device CDS criteria?
The function is not intended to acquire, process or analyze medical images, signals from in vitro diagnostic devices, or patterns or signals from signal acquisition systems. It is intended to display, analyze or print patient or other medical information. It is intended to support or recommend prevention, diagnosis or treatment of a disease or condition to a health care professional. It enables independent review of the recommendations’ basis, with no intended primary reliance on them for an individual patient’s clinical diagnosis or treatment decision.
Is patient-facing decision support software a device?
FDA says functions that support or provide recommendations to patients or caregivers, not health care professionals, meet the device definition. Its existing digital health policies, including enforcement discretion policies, continue to apply to them.
Is an alert for sepsis or stroke non-device CDS?
FDA’s example analyzes patient-specific information to detect stroke or sepsis and alerts a health care professional. It is a device function that fails criteria 3 and 4 because it gives a specific diagnostic output or directive and supports time-critical decisions.
What is enforcement discretion for a single recommendation?
If only one option is clinically appropriate and the function otherwise meets all four criteria, FDA says it intends not to enforce requirements under the Act for that function. The guidance lists variations that fall outside the policy, so each function needs its own check.
Is the Digital Health Policy Navigator a formal determination?
No. FDA says the Navigator's results are not a formal FDA device determination for your product.
Does a 513(g) response clear my software?
No. FDA says 513(g) responses are not device classification decisions and do not constitute clearance or approval for marketing.
Does IHS decide whether my software is a device?
No. Integral Healthcare Solutions documents your organization's own assessment against FDA's guidance. Your regulatory lead and counsel reach the conclusion, and any inquiry to FDA comes from the manufacturer.
Related: compliance services, FDA software premarket pathway support, healthcare AI governance.