A predetermined change control plan (PCCP) describes planned device changes and how they will be assessed. For AI-enabled device software functions, FDA’s August 2025 final guidance recommends a Description of Modifications, Modification Protocol and Impact Assessment. Integral Healthcare Solutions (IHS) drafts these documents for device makers; your regulatory, quality, clinical and data science leads supply evidence and approve them, and your company submits to FDA.
Last updated: November 2026
What is a PCCP for an AI-enabled device software function?
FDA's guidance, Marketing Submission Recommendations for a Predetermined Change Control Plan for Artificial Intelligence-Enabled Device Software Functions, was issued August 18, 2025 and first issued December 4, 2024, under docket FDA-2022-D-2628 (guidance page and PDF, opened November 25, 2026). The guidance is nonbinding; FDA says "should" means suggested, not required.
The legal basis is section 515C of the FD&C Act. For an approved PMA device, no supplemental application is required for a change consistent with an approved PCCP; for a cleared 510(k) device, no new premarket notification is required for a change consistent with an established PCCP (21 U.S.C. 360e-4). The regulations carry the same exceptions, each conditioned on consistency with the plan: 21 CFR 807.81(b)(1)(ii) and 21 CFR 814.39(b).
A PCCP generally covers changes otherwise requiring a new marketing submission. Changes must preserve intended use and, where applicable, substantial equivalence to the predicate. The guidance covers automatic, manual and combined implementation.
What goes into the PCCP document set?
FDA recommends three components, and recommends that a PCCP include only a limited number of modifications that are specific and can be verified and validated.
| Component | What FDA recommends it contains |
|---|---|
| Description of Modifications | Each change’s rationale and resulting specifications; automatic, manual or combined implementation; uniform and/or local application; factors warranting local changes; and update frequency. |
| Modification Protocol | For each modification, four components: data management practices, re-training practices, performance evaluation and update procedures, with pre-defined acceptance criteria. It also states that an unresolvable performance failure is recorded and the modification is not implemented. Testing may be repeated if root cause analysis shows the failure is unrelated to specific aspects of the PCCP. |
| Impact Assessment | Five elements: each modification compared with the unmodified device; benefits and risks, including risks of harm and unintended bias, of each; how the protocol continues to support safety and effectiveness; how one modification affects another; and the cumulative impact of all modifications. |
FDA recommends mapping methods to modifications; a table can do this for multiple changes. Testing should compare the modified device with its original and last modified versions. Test data should be independent of training and tuning data and generally come from multiple sites different from those used for training and tuning (August 2025 guidance, §§IV.B, VII.B–C; page opened November 25, 2026).
For the submission itself, FDA recommends the PCCP be a standalone section with a title and version number, discussed in the cover letter and listed in the table of contents as "Predetermined Change Control Plan." It recommends a labeling statement that the device has an authorized PCCP. FDA's PCCP page says an eSTAR submission can flag a PCCP in the Device Description section.
Who needs a PCCP document set, and what brings the work forward?
A PCCP is an option for device makers planning updates that would otherwise require new marketing submissions (August 2025 guidance, §III; page opened November 25, 2026). Three events can prompt drafting:
- A planned 510(k), De Novo or PMA submission for an AI-enabled software function. A PCCP is established only as part of a marketing authorization.
- A device already on the market that a manufacturer wants to modify to add a PCCP.
- A change to an authorized PCCP. FDA says such changes will generally constitute changes that would otherwise require a new marketing submission.
The volume is growing. FDA's AI-enabled medical devices list listed 1,614 entries when we counted its CSV on November 25, 2026, 335 of them with 2025 decision dates; FDA says the list is not comprehensive. FDA's list of devices with PCCPs listed 240 authorized devices by our count of its table, of which 103 carry 2026 decision dates through September 30, 2026. That list covers all device types.
How does IHS help?
IHS drafts the document set and the procedures around it in this order:
- Gap assessment of your current change records and any draft PCCP against the three components and the four protocol components.
- Drafting the Description of Modifications from your product roadmap, with rationale, implementation mode, update frequency and labeling sections affected.
- Drafting the Modification Protocol procedures for data management, re-training, performance evaluation and updates, with fields for your acceptance criteria and the failure statement.
- Drafting the Impact Assessment framework and the traceability table that ties each modification to its methods.
- Drafting the procedure that connects the PCCP to your quality system: checking a proposed change against the authorized plan, documenting it, routing a change outside the plan, and post-market monitoring.
- Drafting cover-letter language, a version log, the labeling statement and a question list for your regulatory lead's Q-Submission.
- A read of the assembled set against the guidance before you file.
Your company supplies the model and data documentation, validation results, acceptance criteria and statistical plans, the risk file, labeling and your quality system procedures. The limit is plain: IHS is a compliance and program development firm, not a testing laboratory or law firm. IHS does not design or validate the algorithm, set your acceptance criteria, decide whether a modification belongs in a PCCP or needs a new submission, or predict what FDA will do. Your regulatory lead and counsel decide those points, and your company handles every contact with FDA.
What should you have ready?
Checklist sources: Marketing Submission Recommendations for a Predetermined Change Control Plan for Artificial Intelligence-Enabled Device Software Functions (August 18, 2025; sections below) and current 21 CFR 820.10; pages opened November 25, 2026.
- Intended use, which must remain unchanged, and indications for use; certain indication changes may be appropriate (§VI.C).
- Specific, verifiable, validatable modifications and their rationale (§VI).
- Implementation modes, local-change conditions, update frequency and any automatic-update guardrails (§VI.B).
- How new data are collected, annotated, curated, stored, retained, controlled and used for each modification; independent test data (§VII.B(1)).
- Evaluation design, metrics, acceptance criteria, statistical tests and original/last-version comparisons (§VII.B(3)).
- Update, user-communication, labeling and surveillance procedures (§VII.B(4)).
- An Impact Assessment addressing all five elements (§VIII).
- Method-to-modification mapping, optionally a table for multiple modifications (§VII.C).
- Change procedures and, for manufacturers subject to Part 820, a documented quality management system meeting applicable ISO 13485 and other applicable Part 820 requirements, plus other applicable Title 21 requirements as appropriate (21 CFR 820.10(a)–(b)).
- A titled, versioned PCCP section, prominent cover-letter discussion, contents entry and labeling statement (§V.C).
The guidance retains legacy QSR references; your quality lead should map them to current Part 820 under the QMSR. Compare your documents with this checklist on the free introductory call.
How does this compare with other ways to handle device changes?
FDA contrasts a PCCP with a traditional device change, where the manufacturer may need to submit a new or supplemental marketing submission before implementing the change (FDA PCCP page).
| Topic | Traditional change | Authorized PCCP |
|---|---|---|
| When the change is evaluated | After authorization, when the change is planned | During the submission establishing the PCCP, including for an already marketed device |
| Marketing submission for each change | May be needed, depending on the change | Not needed when the change is in the plan and made according to its protocol |
| Main document burden | Per-change analysis; submission when required | Description of Modifications, Modification Protocol and Impact Assessment up front |
Three live IHS pages sit next to this one: the QMSR document control page (21 CFR Part 820), the healthcare AI governance page and the device reporting and complaint handling page. This service focuses on the three-part AI-device PCCP document set. FDA's all-device PCCP draft guidance (August 2024, page) and its draft AI lifecycle guidance (January 2025, page) were still marked draft on the pages we opened.
What does it cost?
The FDA guidance and PCCP pages reviewed do not publish a separate PCCP fee schedule. Fees for the marketing submission itself were not part of this review; verify current fees with FDA. IHS scopes each engagement after a free introductory call.
What this is not
- It is not legal advice. IHS is a consulting firm, not a law firm, and your counsel reads how the statute and regulations apply to your device.
- It is not a prediction or guarantee of any FDA decision on a PCCP, a modification or a device.
- It is not algorithm development or validation testing, and IHS does not file with or correspond with FDA for you.
Frequently asked questions
Regulatory answers below follow FDA’s August 2025 PCCP guidance, §§IV–VII (page opened November 25, 2026).
What is a predetermined change control plan for an AI-enabled device?
It is the documentation, included in an FDA marketing submission, that describes what modifications the manufacturer will make to the device and how each will be assessed. FDA's guidance recommends three parts: a Description of Modifications, a Modification Protocol and an Impact Assessment. After FDA authorizes the PCCP, changes specified in it and implemented according to its protocol need no separate marketing submission.
Does a PCCP let a manufacturer change anything after authorization?
No. FDA recommends specific, verifiable and validatable changes; they must preserve intended use and, for 510(k) devices, substantial equivalence to the predicate. Some changes may be unsuitable for any PCCP. Significant changes outside the authorized plan or its protocol likely need a new marketing submission before implementation.
Which submission types can establish a PCCP?
FDA lists original PMAs; modular PMAs with the PCCP as a review module; 180-day and Panel Track supplements; and Real-Time supplements when the PCCP covers minor device changes and FDA and the manufacturer agree real-time review is feasible. Traditional or abbreviated 510(k)s and original De Novo requests can also establish a PCCP. An affirmative FDA marketing authorization is required.
Can a Pre-Submission or Q-Submission authorize a PCCP?
No. FDA encourages manufacturers to use the Q-Submission Program to get feedback on a proposed PCCP, but FDA does not authorize a PCCP in a Pre-Submission. Your company requests any Q-Submission; Integral Healthcare Solutions can draft the questions for your regulatory lead.
What happens if a manufacturer deviates from its authorized PCCP?
Failure to follow an authorized PCCP is generally a deviation. FDA excludes choosing not to implement a planned change, choosing a new marketing submission instead, and changes that would not require a marketing submission. Other deviations generally cause adulteration and misbranding under sections 501(f)(1) and 502(o); significant changes outside the plan likely need a new submission before implementation.
Does FDA accept PCCPs for models that update automatically?
FDA says it will consider PCCPs for modifications that are implemented automatically, and that this subset adds complexity. It suggests setting boundaries or guardrails that define the range of automatic updates and discussing them through the Q-Submission Program.
What if FDA finds deficiencies in the PCCP during review?
FDA expects review to be interactive and may send a deficiency letter. If deficiencies with the PCCP remain unresolved, FDA may authorize the device upon withdrawal of the PCCP.
Does Integral Healthcare Solutions file anything with FDA or contact FDA for a client?
No. IHS drafts the document set and procedures. Your company reviews them, submits the marketing submission and handles every contact with FDA, including any Q-Submission.